Moniteur patient portable YSPM400V

 
Features:
Body
·8.0 inchhigh-brightness TFT LED
·Support display maximum 8 waveforms
·Support 6 channel ECG waveform display simultaneously 
·Optionaltouch screen
 
Stroage
·Support storage of 168 hours trend table and graph review
·100 minutes waveform review
·500 groups NIBP review
·100 hours of parameter alarm events review and associated parameters
 
Printer
·Built-in high-speed 50 mm thermal printer
 (Brand: SEIKO, Japan)
 
Central System
·Optional:supporting wired or wireless connection to the central monitoring station.
 
Alarm
·Three-level acousto-optic alarm
·Sensor-off alarm
·Paper out alarm
·Support alarm review
·Support alarm pause
 
Operating System
·Multi-display mode to meet different clinical requirement, including standard interface, big font, OxyCRG, trend graph, NIBP review and full leads ECG interface.
·NIBP self-test mode: including overpressure test, static pressure test and air leakage test.
·15 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
·Drug calculationand titration table
·Support multi-language display
·Support online software upgrading by net
 
Interface
Transducer socket
·ECG
·SPO2
·NIBP 
·TEMP 1
·TEMP 2 / Mainstream EtCO2
·Sidestream EtCO2
 
LAN / Central monitor system interface
·The LAN port is for connecting the monitor to a central monitor system network.
 
Input device interface
·Knob (Standard)
·Touch screen (optional)
 
                     
Performance Specifications:
Dimension and Weight
-  Dimension: 175mm*210mm*80mm
-Weight:1.35kg(excluding accessories)
 
Power Supply
-  Adapter:  Output 13-20V DC, 5A
-  Voltage: AC100~240V, 50/60HZ

Display
-  8" color TFT LED 
-  Resolution: 800*600 pixels
 
Battery (Pluggable)
-  Type: Rechargeable lithium battery 14.8V/2200mAh
-  Charge Cycle: ≥500 times
-  Working time: 5.5 hours
 
Recorder (Option)
-  Method: Thermal printer
-  Paper width: 50 mm (1.97 in)
-  Printing speed: 25/ 50 mm/s
-  Trace: Max. 2 tracks
-  Recording way: Real-time Recording, 
                              Alarm Recording
 
Alarm
-  Level: Low, medium and high
-  Indication: Auditory and visual
-  Alarm volume adjustable
-  Alarm pause time: 1min, 2min
-  Parameter alarm type: Latch/ Unlatch
 
Input Device
-  Touch screen (option)
-  Knob (Standard)
-  Keypad input (Standard)
 
System Output & Extensible Interface
-  Ethernet Network: 1 standard RJ45 socket
 
Operating Environment
-  Temperature:  5 ~ 40 ℃
-  Humidity:  15% ~ 80% (non-condensing)
-  Atmosphere pressure:  86 KPa ~ 110 KPa

Transportation and Storage Environment
-  Temperature:  -20~50 ℃
-  Humidity:  10%~90%  (non-condensing)
-  Atmosphere pressure:  86 KPa ~ 110 KPa
 
Safety
-  IEC60601-1 Approved, CE marking according to MDD93/42/EEC
-  With reference to RoHS Directive 2011/65/EU recasting
 
Trend & Reviewing
-  Trend: 168 hours
-  NIBP measurement reviewing: 500 groups
-  Waveform review: 100 minutes
-  Alarm event: 100hours of parameter alarms events and associated parameter
 
ECG
-  Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
-  Gain: 2.5mm/mV(×0.25), 5.0mm/mV(×0.5), 10mm/mV (×1), 20mm/mV (×2),
-  Sweep speed: 6.25 mm/s12.5 mm/s, 25 mm/s, 50 mm/s
-  ECG signal range:  ±6 mV
-  Accuracy: ±1bpm/ ±1%, whichever is greater
-  Resolution: 1 bpm
-  Leakage Current < 10 μA
-  Input impedance: > 5.0Mohm
-  CMRR: Diagnostic Mode >90 dB;
       Monitor Mode>100dB
       Surgery Mode>100dB
-  Baseline Recovery: ≤ 5s after defibrillation (Monitor or Surgery mode)
-  Bandwidth: Surgery 1 ~ 20Hz
   Monitor 0.5 ~ 40 Hz
   Diagnostic 0.05 ~ 130 Hz
-  Against electrosurgical interference and defibrillation
-  Indication of Electrode Separation: Every electrode (exclusive of RL)
 
Heart Rate
-  Measure range:
    Big animal 0bpm; 15 ~ 300 bpm
    Small and middle: 0bpm; 15 ~ 450 bpm
-  Resolution: 1 bpm
-  Accuracy: ± 1% or ± 1bpm, whichever is greater

ST Measurement
-  Range: -2.0 ~ +2.0 mV
-  Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is greater
-  Other range: unspecified
 
Respiration
-  Method: Impedance between RA-LL,
-  Gain: ×0.25, ×0.50, ×1, ×2,
-  Respiration Rate: 0bpm, 0 ~ 150 BrPM
-  Sweep speed:  6.25 mm/s, 12.5 mm/s, 25mm/s
-  Resolution: 1 BrPM
-  Accuracy: ±2 bpm or ±2%, whichever is greater
-  Apnea Alarm: 10 ~ 40 s
 
NIBP
-  Method: Oscillometric
-  Measure mode: Manual, Auto, STAT
-  Measure Interval in AUTO Mode 1~480 min
-  STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
-  Measure and Alarm Range:
Big animal: SYS:  40 ~ 270 mmHg
                    DIA:  10 ~ 215mmHg    
                    MEAN:  20 ~ 235 mmHg
Middle animal: SYS:  40 ~ 200 mmHg
                          DIA:  10 ~ 150mmHg    
                          MEAN:  20 ~ 165 mmHg
Small animal: SYS:  40 ~ 135 mmHg
                        DIA:  10 ~ 100mmHg    
                        MEAN:  20 ~ 110 mmHg
-  Static pressure accuracy: ±3mmHg
-  Resolution: 1mmHg
-  Accuracy: Maximum Mean error ±5mmHg          
         Maximum Standard deviation ≤8mmHg
-  Over pressure Protection: Dual protection via software & hardware
 
Temperature
-  Technique: Thermistor probe (2.25K)
-  Channel:  Dual-channel, provide T1; T2; ΔT
-  Measuring and Alarm Range: 0.0 ℃~ 50 ℃ (32℉~ 122℉)
-  Unit: Celsius (℃), Fahrenheit ( ℉)
-  Resolution: 0.1℃or 1℉
-  Accuracy: ±0.1℃(exclusive probe)

Performance Specifications:
SpO2 (Analog Technic)
-  Measurement Range: 0 ~ 100 %
-  Resolution: 1 %
-  Accuracy:  ±2% (70% ~ 100%)
Unspecified (0 ~ 69%)
 
Pulse Rate
-  Measuring and Alarm Range: 20~450bmp
-  Accuracy: ±3%
-  Resolution: 1bpm
 
EtCO2 (Mainstream/ Sidestream) (Option)
-  Measure method:  Non-dispersive infrared (NDIR)
-  Measure Range:  0 ~19.7% (0 ~ 150 mmHg)
                 0 ~ 20 kPa
-  Resolution: 0.1 mmHg
-  CO2 Accuracy: 0 ~ 40 mmHg, ±2 mmHg 41 ~ 70 mmHg, ±5% of reading  71 ~ 100 mmHg, ±8% of reading 101~ 150 mmHg, ±10% of reading at 760 mmHg, ambient temperature of 25℃)
-  Respiratory Rate:  Range: 3 ~150 BrPM Accuracy: ±1 bpm
 
EtCO2 (Micro-stream) (Option)
-  Measure method:  Non-dispersive infrared (NDIR)
-  Measure Range:  0 ~19.7% (0 ~150 mmHg)
                 0 ~ 20 kPa
Obtenir le prix
Notez votre demande ci-dessous et nous vous contacterons sous 24h!

YOUR MAY ALSO LIKE ...
Moniteur de patient vétérinaire YSPM500V
Moniteur de patient vétérinaire YSPM500V
Moniteur de patient vétérinaire YSPM510V
Moniteur de patient vétérinaire YSPM510V
Moniteur de pression artérielle vétérinaire YSBP80V
Moniteur de pression artérielle vétérinaire YSBP80V
Tensiomètre vétérinaire YSDBP100V
Tensiomètre vétérinaire YSDBP100V