We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.

Professional Medical Equipment Supplier - Yuesen Med Professional Medical Equipment Supplier - Yuesen Med

Hot Products

YSX500D 50kW DR system set up and put into service in Cambodia.

YSENMED YSX500D 50kW digital x-ray system has been successfully set up and put into service in a hospital in Cambodia.

YSX056-PE serving as a vehicle-mounted x-ray in the Philippines

YSX056-PE 5.6kW portable x-ray unit has been adapted to fit on a truck, to provide mobile x-ray examination service for remote communities in the Philippines.

X Ray Machine  To  Zimbabwe

x ray machine, 50KW x ray machine

Microscope To Malawi

Achromatic objectives: 4X、10X、40X(S), 100X(S、Oil) Wide field eyepiece: WF10X(WF16X for option) Eyepiece head: Sliding binocular head inclined at 45° Stage: Double layer mechanical stage size 140X140mm, moving range 75X45mm Focusing: Coaxial coarse and

GE Mobile X-ray machine fault

Views : 1732
Author : Monica Liang
Update time : 2015-12-21 08:00:00

The machine over-current protection contactor triped and system was automatic power off when open the machine and pressed positioning lamp or exposure preparation in use.

Failure analysis and maintenance:

AMX-4+ system is consist of power box, charging control, CPU board, high voltage generator, motor control and remote control receiving. Every model is controled and detected by CPU. Then check attached machine circuit diagram and consider that positioning lamp(AC 24V, 150W) and exposure preparation(Rotating anode start ) are provided power by T1 transformer, it is controled by 46-232786G2.

The next step you could take down the board and measure, you would find that F93, which had insurance tube, is burned off and effective tube was breakdown. After replacing the same model tube, connect power and open the machine, fault remains. Then you would doubt CPU control singal abnormal and increase the load spurious. It leads to excessive current and power tube punctured. The following you could use digital multimeter to measure and 60Hz control singal is normal.

Replace the tube again and then connect power, the two test points are 5V and 6V. This two points are not balance. After power off, take down promote tube Q61, there is leakage phenomenon. After replacing the same model tube, connect power and measure two points are 6.0V. After opening the machine and self-inspection, fault is solved and x-ray machine can work normally.

Related News
Read More >>
Blink and You'll Miss It: How a 0.298 sec/rev Scanner Makes Cardiac CT Feel Like Slow Motion Blink and You'll Miss It: How a 0.298 sec/rev Scanner Makes Cardiac CT Feel Like Slow Motion
Dec .27.2025
Explore how the 0.298 sec/rev Cardiac CT Scanner revolutionizes cardiac imaging with speed and precision.
From 40–150kV to 800mA Bursts: How a 65KW/800mA DRF System Works Like a Turbo Engine in Medical Imaging From 40–150kV to 800mA Bursts: How a 65KW/800mA DRF System Works Like a Turbo Engine in Medical Imaging
Dec .21.2025
In this article, we'll explore how the 65KW/800mA Digital Radiography and Fluoroscopy System (DRF) operates, its components, and why it's revolutionizing the field of medical imaging.
Small Spot, Big Impact: How Dual Focal Spots (0.2/0.4mm) in a Digital Mammography X-ray System Sharpen Every Detail Small Spot, Big Impact: How Dual Focal Spots (0.2/0.4mm) in a Digital Mammography X-ray System Sharpen Every Detail
Dec .20.2025
When it comes to breast cancer detection, precision is everything. That's where the Digital Mammography X-ray System steps in, particularly with its innovative dual focal spots.
Power Up Your Practice: The Benefits of the 15kW Digital C-arm X-ray System Power Up Your Practice: The Benefits of the 15kW Digital C-arm X-ray System
Dec .15.2025
In this article, we'll explore the myriad benefits of the 15kW Flat Panel Detector Digital C-arm X-ray System, the YSX-C715 model, how it can enhance your practice, and why it's worth considering for your facility.