We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.

Professional Medical Equipment Supplier - Yuesen Med Professional Medical Equipment Supplier - Yuesen Med

Hot Products

YSX500D 50kW DR system set up and put into service in Cambodia.

YSENMED YSX500D 50kW digital x-ray system has been successfully set up and put into service in a hospital in Cambodia.

YSX056-PE serving as a vehicle-mounted x-ray in the Philippines

YSX056-PE 5.6kW portable x-ray unit has been adapted to fit on a truck, to provide mobile x-ray examination service for remote communities in the Philippines.

X Ray Machine  To  Zimbabwe

x ray machine, 50KW x ray machine

Microscope To Malawi

Achromatic objectives: 4X、10X、40X(S), 100X(S、Oil) Wide field eyepiece: WF10X(WF16X for option) Eyepiece head: Sliding binocular head inclined at 45° Stage: Double layer mechanical stage size 140X140mm, moving range 75X45mm Focusing: Coaxial coarse and

GE Mobile X-ray machine fault

Views : 2362
Author : Monica Liang
Update time : 2015-12-21 08:00:00

The machine over-current protection contactor triped and system was automatic power off when open the machine and pressed positioning lamp or exposure preparation in use.

Failure analysis and maintenance:

AMX-4+ system is consist of power box, charging control, CPU board, high voltage generator, motor control and remote control receiving. Every model is controled and detected by CPU. Then check attached machine circuit diagram and consider that positioning lamp(AC 24V, 150W) and exposure preparation(Rotating anode start ) are provided power by T1 transformer, it is controled by 46-232786G2.

The next step you could take down the board and measure, you would find that F93, which had insurance tube, is burned off and effective tube was breakdown. After replacing the same model tube, connect power and open the machine, fault remains. Then you would doubt CPU control singal abnormal and increase the load spurious. It leads to excessive current and power tube punctured. The following you could use digital multimeter to measure and 60Hz control singal is normal.

Replace the tube again and then connect power, the two test points are 5V and 6V. This two points are not balance. After power off, take down promote tube Q61, there is leakage phenomenon. After replacing the same model tube, connect power and measure two points are 6.0V. After opening the machine and self-inspection, fault is solved and x-ray machine can work normally.

Related News
Read More >>
Saudi Distributor Purchases the 5th YSENMED YSX056-PF VET Portable Veterinary DR with Clear Imaging and Reliable Support Saudi Distributor Purchases the 5th YSENMED YSX056-PF VET Portable Veterinary DR with Clear Imaging and Reliable Support
Jul .06.2026
A Saudi distributor purchased the fifth YSENMED YSX056-PF VET 5.6kW Veterinary Portable X-ray Unit, praising its sharp image quality, prompt remote support, and expressing plans for future repeat orders.
Successful Installation of YSOT-YT5D Surgical Table & VG70 Ventilator in Ghana Hospital Successful Installation of YSOT-YT5D Surgical Table & VG70 Ventilator in Ghana Hospital
Jul .05.2026
YSENMED successfully delivered the YSOT-YT5D 5-Function Electric Surgical Table and VG70 Trolley Turbine Ventilator to a hospital in Ghana. Both units were installed successfully and are now ready to support daily clinical procedures.
Cameroon Hospital Expands Imaging Capacity with YSENMED 32kW Digital X-Ray Syste Cameroon Hospital Expands Imaging Capacity with YSENMED 32kW Digital X-Ray Syste
Jul .04.2026
Distributor Hallie reordered the YSENMED 32kW Digital X-Ray Systems for a hospital in Cameroon. The customer praised the excellent image quality and expressed satisfaction with YSENMED's professional after-sales support.
Veterinary Distributor in the Philippines Chooses YSENMED YSDR-VET320 for High-Quality Imaging in Clinical Practice Veterinary Distributor in the Philippines Chooses YSENMED YSDR-VET320 for High-Quality Imaging in Clinical Practice
Jul .03.2026
After installation and use, Dr. Michael expressed strong satisfaction with the system, especially praising its clear and high-resolution imaging quality.