We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.

Professional Medical Equipment Supplier - Yuesen Med Professional Medical Equipment Supplier - Yuesen Med

Hot Products

YSX500D 50kW DR system set up and put into service in Cambodia.

YSENMED YSX500D 50kW digital x-ray system has been successfully set up and put into service in a hospital in Cambodia.

YSX056-PE serving as a vehicle-mounted x-ray in the Philippines

YSX056-PE 5.6kW portable x-ray unit has been adapted to fit on a truck, to provide mobile x-ray examination service for remote communities in the Philippines.

X Ray Machine  To  Zimbabwe

x ray machine, 50KW x ray machine

Microscope To Malawi

Achromatic objectives: 4X、10X、40X(S), 100X(S、Oil) Wide field eyepiece: WF10X(WF16X for option) Eyepiece head: Sliding binocular head inclined at 45° Stage: Double layer mechanical stage size 140X140mm, moving range 75X45mm Focusing: Coaxial coarse and

The several faults of Ultrasound probe

Views : 1896
Author : Monica Liang
Update time : 2015-12-21 08:00:00

Ultrasound diagnosis equipment uses ultrasonic to extract all kinds of information of human tissue, such as organization interface information, blood flow information, so ultrasound probe, which can transfer information. Ultrasound probe is the key part of ultrasound diagnosis and its value has accounted for a quarter to a third of one ultrasound instrument total value, like B ultrasound probe is the same. In addition, doctors use it to check patients and it is easy to have faults. Therefore, it is necessary for reasonable using and maintenance.

In general, there are several faults of ultrasound probe as follows: Appearance cracking, internal sparkling, sensitivity decreasing, cable cracking and electrode lead break.

 First, Appearance cracking, internal sparkling, reduced sensitivity.

(1)Although ultrasound probe manufacturers are different, the basic structure is the same in general. Put the transducer in probe cavity of plastic wrap.

 As we know, when we examinate, checking parts have to been embrocated enough lotus root mixture, but this lotus root mixture is the biggest threat of plastic wrap. In most conditions, probe surface lotus root mixture is not clear and cause appearance cracking after finishing operation. After a long time, lotus root mixture has corrosion for the plastic and rubber band and damage probe. Therefore we must clean the surface lotus root mixture after finishing work and you had better use clean and wet soft cloth to wipe it; The other reason of appearance cracking is that probe is often influenced by the sun and large temperature difference. Because the sun will make plastic materials early ageing and deformation. So, ultrasound equipment should be placed in constant greenhouse and avoid light ray. For example probe frequency instability, you can consider probe moisture, then there is no effect after using no water copper sulfate moisture absorption; The following you use special tool to open probe shell and take down the circuit board and find that there is more circuit board oil and use 99.5% pure alcohol to soak ten minutes; Then use fuzz wipe oil and dirty; The next step, you can use fan to blow dry and put it back and open the machine, probe fault is removed.

(2) Internal sparkling. In general, probe production process defect causes sensitivity decrasing. It will happened after a period of time, especially bad penetration ability after adjusting host parameters.

Second, Circle image causes bad local

In general, chip problem cause this fault. For example, there is a white bright field from up to down and there is no imge information in this field. After checking, the probe internal chip damage causes this.

Third, Cable craze or electrode lead break

Due to probe move back and forth in use process, it will pull cable automatically, especially in turning probe process, it will make probe and the cable wind. Therefore, bad operation habit or cable placed bad cause this fault. For this fault, you can open probe shell and rearrange all lines and weld well, then use super glue sticky shell well.

 
Related News
Read More >>
Seychelles Clinic Successfully Equips with YSENMED Ultrasound, Lab & Patient Care Devices Seychelles Clinic Successfully Equips with YSENMED Ultrasound, Lab & Patient Care Devices
Jul .07.2026
Ribail Babie from Seychelles equipped a private clinic with YSENMED ultrasound, hematology analyzer, biochemistry analyzer, EMG system, patient monitor and more. The customer is highly satisfied with both the equipment performance and our professional ser
Saudi Distributor Purchases the 5th YSENMED YSX056-PF VET Portable Veterinary DR with Clear Imaging and Reliable Support Saudi Distributor Purchases the 5th YSENMED YSX056-PF VET Portable Veterinary DR with Clear Imaging and Reliable Support
Jul .06.2026
A Saudi distributor purchased the fifth YSENMED YSX056-PF VET 5.6kW Veterinary Portable X-ray Unit, praising its sharp image quality, prompt remote support, and expressing plans for future repeat orders.
Successful Installation of YSOT-YT5D Surgical Table & VG70 Ventilator in Ghana Hospital Successful Installation of YSOT-YT5D Surgical Table & VG70 Ventilator in Ghana Hospital
Jul .05.2026
YSENMED successfully delivered the YSOT-YT5D 5-Function Electric Surgical Table and VG70 Trolley Turbine Ventilator to a hospital in Ghana. Both units were installed successfully and are now ready to support daily clinical procedures.
Cameroon Hospital Expands Imaging Capacity with YSENMED 32kW Digital X-Ray Syste Cameroon Hospital Expands Imaging Capacity with YSENMED 32kW Digital X-Ray Syste
Jul .04.2026
Distributor Hallie reordered the YSENMED 32kW Digital X-Ray Systems for a hospital in Cameroon. The customer praised the excellent image quality and expressed satisfaction with YSENMED's professional after-sales support.