Item No.: YSAV201PV
High Performance Portable Vet Ventilator YSAV201PV Hot
| Product parameters | |
| Brand: | YSENMED |
| MOQ: | 1 |
| Package: | Carton Package |
| Ship From: | Guangzhou |
| Lead Time: | 7-15 days |
| Customization: | YES |
Description
Portable Veterinary Ventilator / Respirator

Vet Ventilator Specification

| Display mode | High-definition 5.7”LCD screen display | |
| Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH | |
| Ventilation parameters | Tidal volume | 50~1500ml |
| Rate | 2~99bpm | |
| SIMV rate | 2~20bpm | |
| I:E | 2:1~1:8 | |
| Inspiratory trigger pressure | -1.0~2.0 kPa | |
| PEEP | 0~2.0 kPa | |
| Pressure range | 1.0~6.0 kPa | |
| SIGH | 1.5 times the inspiratory time | |
| Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure | |
| Security Alarm System | ||
| Airway pressure alarm | upper limit setting range | 1.0~6.0kPa |
| low limit setting range | 0.4~2.0kPa | |
| Per-minute ventilation volume alarm | upper limit setting range | 3.0~30L/min |
| low limit setting range | 1.0~10L/min | |
| Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa | |
| Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds | |
| Power alarm | ||
| Power | AC 220V±10% 50Hz±1Hz UPS, Storage battery |
|